EVN feedback tool

Feedback for EG037A

Friend Site Result Problem Text
TaoAn Shanghai Success None None
Zhang hua Urumqi Success None The compressor of C-band doesn't working.we used room temperature receiver. Otherwise no known problems.
andrea orlati at S Medicina Success None The vsn snd the front label of the second disk pack do not match: MPI-0053 in the front, jb-0002 reported in the log file. A part from this, no known problems.
Michael Lindqvist at S Onsala Success None No known problems.
P. Cassaro Noto Minor failures FS From 22:26 to 22:50 UT scan lost due to FS problems.
Kaz Borkowski Torun Success None Four minutes beginning at 17:31 UT lost due to a problem with control of telescope elevation drive. No other known problems.
A. Polatidis Effelsberg Minor failures RECE VC05 still not working. Observing system restarted at 2150 UT. There may be loss of data before this.
Tony Foley Westerbork Failure RECO FS Wb used in phased array mode. Big problems with the Westerbork file server (RAID card and possible also disk damage); we had to replace the server on Friday evening 1st June. The file server had also given errors on Thursday. We recorded from 15:30 to 16:44 (operator error had not enabled 2nd bank while checking the local system). Observations restarted at 20:49UT, but most of the data from 20:49 to 00:30 gave large numbers of data_check errors.
Tony Foley Westerbork Minor failures None Single dish was used. Please change the coordinates in the vexfile accordingly. Supplemental : Scan_check gave large errors in the restarted scan at 21:03. scan checks from 16:44 to 21:03 are looking at the same scan.