| Friend | Site | Result | Problem | Text |
|---|---|---|---|---|
| scalambra | Medicina | Minor failures | None | None |
| Peng Li | Urumqi | Success | None | None |
| Michael Lindqvist | Onsala | Success | None | No known problems. |
| Jonathan Quick | Hartebeesthoek | Minor failures | RECE TSYS | The RCP channel of the IF distributor has a sticky programmable attenuator causing dropouts after Tsys measurements affecting multiple scans from 02:37-02:44, 02:48-02:49, 02:59-03:01, 05:10, 06:12-06:13, 06:22-06:27, 06:43-06:46, 07:03-07:06, 07:13-07:14, 07:34-07:43, 07:54, 08:04-08:06, 08:14-08:21, 08:35-08:37. No other known problems. |
| K. Borkowski | Torun | Minor failures | RECO FS | Two breaks occurred. One was due to a telescope control problem at 2:42 (probably data recorded off source for most of 35 min that followed), and the other at 4:05 UT, when Mark5A hanged. Data were completely lost (not recorded) for 1 hour and 15 min. |
| Lysakova | Badary | Minor failures | ANT | Scans no0138,0139,0140 failed due to antenna problems. |
| Kiselev P. Kvashnin E. Garkusha A. | Zelenchukskaya | Success | None | None |
| Uwe Bach | Effelsberg | Success | None | None |
| Geert Kuper | Westerbork | Success | None | None |
| Bo Xia | Shanghai | Success | None | From UT 6:43 to 6:54,mk5b syncerr error,there was some probles with our VSI-C card.Reset the VSI-C card. |
| Alastair Gunn | Jodrell_Bank | Success | None | No known problems. |